Equipment Checklist for AHTD 455
Water Sensitivity Test for Compacted Bituminous Mixtures

 
All equipment shall be inspected for compliance with appropriate dimension s, as stated in each standard, as well as to determine if it is in satisfactory operation condition (i.e. clean, tight and serviceable). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.

 



Item Vacuum Chamber
Description Capable of withstanding partial vacuums lower than 30 mm Hg absolute pressure, of appropriate size and such that water has access to all sides of the specimen
   
Calibration Item Check physical condition
Calibration Procedure Visual
Calibration Interval 12 months


Item Vacuum System
Description Vacuum pump or aspirator, and a manometer or pressure gauge for measuring the vacuum
   
Calibration Item Verify proper operation
Calibration Procedure Measure pressure
Calibration Interval 12 months


Item Water Bath
Description Constant temperature, able to maintain a temperature of 60 ± 1 C and of sufficient capacity to hold fifteen specimens
   
Calibration Item Verify temperature readings; check water level
Calibration Procedure Verify with calibrated thermometer
Calibration Interval Daily or prior to use


Item Oven
Description Capable of heating to 350 F
   
Calibration Item Verify temperature readings
Calibration Procedure Verify with calibrated thermometer
Calibration Interval 4 months


Item Balance
Description Readable to 0.1% of sample mass or better
   
Calibration Item Verify mass readings
Calibration Procedure AASHTO M 231
Calibration Interval 12 months


Item Load measurement system with recording device
Description Must have breaking head and loading jack, as per AASHTO T 245
   
Calibration Item Verify load measurements
Calibration Procedure Measure loads
Calibration Interval 12 months


   
Reference the following specifications
AASHTO T 245 Resistance to Plastic Flow of Bituminous Mixtures Using Marshall Apparatus