| Item
| Balance
|
|
| Description
| Readable to 0.1% of sample mass or better
|
|
|
|
|
|
| Calibration Item
| Verify mass readings
|
|
| Calibration Procedure
| AASHTO M 231, Class G5
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Sieves
|
|
| Description
| #4, #8, or other as needed
|
|
|
|
|
|
| Calibration Item
| Check openings and physical condition
|
|
| Calibration Procedure
| AASHTO M 92
|
|
| Calibration Interval
| 6 months
|
|
|
|
|
|
| Item
| Sample container
|
|
| Description
| Wire basket of 3.35 mm (No. 6) or finer mesh, or appropriate
bucket
|
|
|
|
|
|
| Calibration Item
| Check physical condition
|
|
| Calibration Procedure
| Visual
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Oven
|
|
| Description
| Must be maintained at 110 ± 5 C
|
|
|
|
|
|
| Calibration Item
| Verify temperature
|
|
| Calibration Procedure
| Verify with calibrated thermometer
|
|
| Calibration Interval
| 4 months
|
|
|
|
|
|
| Item
| Suspension Apparatus
|
|
| Description
| Wire suspending container should be of smallest practical
size
|
|
|
| Water level must completely cover the container
|
|
|
|
|
|
| Calibration Item
| Check Physical Condition
|
|
| Calibration Procedure
| Visual
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Water Bath
|
|
| Description
| Maintained at a constant temperature of 23 ± 1.7 C and
with a constant water level
|
|
|
|
|
|
| Calibration Item
| Verify temperature; Check water level
|
|
| Calibration Procedure
| Verify with calibrated thermometer
|
|
| Calibration Interval
| Daily or Prior to Use
|
|
|
|
|
|
| Item
| Thermometer
|
|
| Description
| Capable of accurately measuring temperatures used in this
test method
|
|
|
|
|
|
| Calibration Item
| Verify temperature readings
|
|
| Calibration Procedure
| ASTM E 1, ASTM E 77
|
|
| Calibration Interval
| 6 months
|
|
|
|
|
|
| Item
| Incidental Equipment
|
|
| Description
| Pans, spoon, hairdryer (optional)
|
|
|
|
|
|
| Calibration Item
| Check physical condition
|
|
| Calibration Procedure
| Visual
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
|
|
|
|
|
| Reference the following specifications
|
|
|
|
|
|
| AASHTO T 2
| Standard Practice for Sampling Aggregates
|
|
| AASHTO T 27
| Sieve Analysis of Fine and Coarse Aggregates
|
|
| AASHTO T 84
| Specific Gravity and Absorption of Fine Aggregate
|
|
| AASHTO T 248
| Reducing Samples of Aggregate to Testing Size
|
|
| AASHTO T 255
| Total Moisture Content of Aggregate by Drying
|
|
|
|
|