| Item
| Balance
|
|
| Description
| Readable to 0.1% of sample mass or better
|
|
|
|
|
|
| Calibration Item
| Verify mass readings
|
|
| Calibration Procedure
| AASHTO M 231
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Sieves
|
|
| Description
| As required for desired application
|
|
|
|
|
|
| Calibration Item
| Check openings and physical condition
|
|
| Calibration Procedure
| AASHTO M 92
|
|
| Calibration Interval
| 6 months
|
|
|
|
|
|
| Item
| Mechanical Sieve Shaker
|
|
| Description
| Must have both lateral and vertical motion
|
|
|
|
|
|
| Calibration Item
| Check sieving thoroughness
|
|
| Calibration Procedure
| AASHTO T 30 Paragraph 6.7
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Oven
|
|
| Description
| Must be maintained at 110 +/- 5 C
|
|
|
|
|
|
| Calibration Item
| Verify temperature
|
|
| Calibration Procedure
| Verify with calibrated thermometer
|
|
| Calibration Interval
| 4 months
|
|
|
|
|
|
| Item
| Incidental Equipment
|
|
| Description
| Containers, filter paper, etc.
|
|
|
|
|
|
| Calibration Item
| Check physical condition
|
|
| Calibration Procedure
| Visual
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Thermometer
|
|
| Description
| Capable of accurately measuring temperatures used in this
test method
|
|
|
|
|
|
| Calibration Item
| Verify temperature readings
|
|
| Calibration Procedure
| ASTM E 1, ASTM E 77
|
|
| Calibration Interval
| 6 months
|
|
|
|
|