| Item
| Balance
|
|
| Description
| Readable to 0.1% of sample mass or better
|
|
|
|
|
|
| Calibration Item
| Verify mass readings
|
|
| Calibration Procedure
| AASHTO M 231
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
| Item
| Oven or other approved heat source
|
|
| Description
| Must be maintained at 110 ± 5 C
|
|
|
|
|
|
| Calibration Item
| Verify temperature
|
|
| Calibration Procedure
| Verify with calibrated thermometer
|
|
| Calibration Interval
| 4 months
|
|
|
|
|
|
| Item
| Incidental Equipment
|
|
| Description
| Sample container, stirrer (spoon)
|
|
|
|
|
|
| Calibration Item
| Check physical condition
|
|
| Calibration Procedure
| Visual
|
|
| Calibration Interval
| 12 months
|
|
|
|
|
|
|
|
|
|
|
| Reference the following specifications
|
|
|
|
|
|
| AASHTO T 2
| Standard Practice for Sampling Aggregates
|
|
| AASHTO T 84
| Specific Gravity and Absorption of Fine Aggregate
|
|
| AASHTO T 85
| Specific Gravity and Absorption of Fine Aggregate
|
|
|
|
|