Equipment Checklist for AASHTO T 231
Capping Cylindrical Concrete Specimens

 
All equipment shall be inspected for compliance with appropriate dimensions, as stated in each standard, as well as to determine if it is in satisfactory operating condition (i.e.; adherence of mortar, deformed molds, etc.). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.

 

 

  Item Capping plates
  Description Glass, metal, polished granite, or stone used for forming sulfur caps against
     
  Calibration Item Gouges, groves, indentations
  Calibration Procedure AASHTO T 231 Sec. 4.1
Visual inspection
  Calibration Interval None stated
 

  Item Alignment devices
  Description Guides to insure that a cap does not depart from perpendicularity
     
  Calibration Item See the general statement above
  Calibration Procedure  
  Calibration Interval  
 

  Item Melting pots for sulfur mortar
  Description Metal pot, or a pot lined with material that is nonreactive with sulfur, used to melt sulfar for making caps
     
  Calibration Item See the general statement above
  Calibration Procedure  
  Calibration Interval  
 

  Item Capping material
  Description High-strength gypsum cement or sulfur mortar used to cap the ends of cylinders
     
  Calibration Item Strength
  Calibration Procedure AASHTO T 231 Sec. 5.2
AASHTO T 231 Sec. 5.2.2
  Calibration Interval Test new lots of material and at intervals not to exceed 3 months
 

  Item Exhaust hood
  Description Used to exhaust gases from the melting of sulfur
     
  Calibration Item See the general statement above
  Calibration Procedure  
  Calibration Interval  
 

  Item Thermometer
  Description Used to monitor the temperature of the melting sulfur
     
  Calibration Item See the general statement above
  Calibration Procedure  
  Calibration Interval  
 

  Item Hardened cap
  Description Check that the caps are remaining plane throughout the day
     
  Calibration Item Cap planeness
  Calibration Procedure AASHTO T 231 Sec. 6.2.4
  Calibration Interval Check 3 specimens each day