Equipment Checklist for AASHTO T 166 - 93
Bulk Specific Gravity of Compacted Bituminous Mixtures Using Saturated Surface-Dry Specimens

 
All equipment shall be inspected for compliance with appropriate dimension s, as stated in each standard, as well as to determine if it is in satisfactory operation condition (i.e. clean, tight and serviceable). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.


METHOD A



Item Balance
Description Readable to 0.1% of sample mass or better, with weigh-below capability
   
Calibration Item Verify mass readings
Calibration Procedure AASHTO M 231
Calibration Interval 12 months


Item Suspension Apparatus
Description Wire suspending container should be of smallest practical size
Water level must completely cover the container
   
Calibration Item Check Physical Condition
Calibration Procedure Visual
Calibration Interval 12 months


Item Water Bath
Description Maintained at a constant temperature of 25 ± 1 C and with a constant water level
   
Calibration Item Verify temperature; Check water level
Calibration Procedure Verify with calibrated thermometer
Calibration Interval Daily or Prior to Use


Item Thermometer
Description Capable of accurately measuring temperatures used in test method
   
Calibration Item Verify temperature readings
Calibration Procedure ASTM E 1, ASTM E 77
Calibration Interval 6 months


Item Oven
Description Capable of maintaining temperature of 52 ± 3 C
(125 ± 5 F)
   
Calibration Item Verify temperature
Calibration Procedure Verify with calibrated thermometer
Calibration Interval 4 months


Item Incidental Equipment
Description Absorbent towels, timer, pan (if using Method C)
   
Calibration Item Check physical condition
Calibration Procedure Visual
Calibration Interval 12 months


 

METHOD B: Equipment in lieu of, or in addition to, that required for Method A



Item Water Bath (Method B)
Description Maintained at a constant temperature of 25 ± 0.5 C and with a constant water level
   
Calibration Item Verify temperature; Check water level
Calibration Procedure Verify with calibrated thermometer
Calibration Interval Daily or Prior to Use


Item Thermometer (Method B)
Description Range of 19 C to 27 C, graduated in 0.1 C subdivisions
   
Calibration Item Verify temperature readings
Calibration Procedure ASTM E 1, ASTM E 77
Calibration Interval 6 months


Item Volumeter (Method B)
Description Appropriate capacity, having tapered lid and capillary bore
   
Calibration Item Check physical condition
Calibration Procedure Visual
Calibration Interval 12 months


 

METHOD C: Equipment in lieu of, or in addition to, that required for Method A



Item Oven (Method C)
Description Maintained at a constant temperature of 110 ± 5 C
   
Calibration Item Verify temperature;
Calibration Procedure Verify with calibrated thermometer
Calibration Interval 4 months