Equipment Checklist for AASHTO T 152
Air Content by the Pressure Method

 
All equipment shall be inspected for compliance with appropriate dimensions, as stated in each standard, as well as to determine if it is in satisfactory operating condition (i.e.; adherence of mortar, deformed molds, etc.). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.

 

 

  Item Type A meter
  Description Measuring bowl and cover assembly made of steel, hard metal, or other hard material
     
  Calibration Item Volume of bowl
  Calibration Procedure AASHTO T 152 Sec. Annex A1, A1.3
  Calibration Interval Occasionally
     
  Calibration Item Expansion factor
  Calibration Procedure AASHTO T 152 Sec. Annex A1.5.1
  Calibration Interval none stated
     
  Calibration Item Calibration reading
  Calibration Procedure AASHTO T 152 Sec. Annex A1.6
  Calibration Interval non stated
     
  Calibration Item Operating pressure
  Calibration Procedure AASHTO T 152 Sec. Annex A1.7, A1.8
  Calibration Interval Frequently
     
    All of the above calibration items shall be checked after rough handling, changes in barometric pressure, changes of more than 600 ft. in elevation
 

  Item Spray tube for a Type A meter
  Description Used to aid the addition of water to a meter
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Type B Meter
  Description Measuring bowl and cover assembly made of steel, hard metal, or other hard material
     
  Calibration Item Volume of Bowl
  Calibration Procedure AASHTO T 152 Sec. Annex A1.3
  Calibration Interval Occasionally
     
  Calibration Item Expansion factor
  Calibration Procedure AASHTO T 152 Sec. Annex A1.5.2
  Calibration Interval none stated
     
  Calibration Item Graduation on the dial face
  Calibration Procedure AASHTO T 152 Sec. Annex A1.9
  Calibration Interval Frequently
     
    All of the above calibration items shall be checked after rough handling
 

  Item Calibration vessel
  Description A vessel of known volume that is used to calibrate a Type B meter
     
  Calibration Item Weight of water required to fill the vessel
  Calibration Procedure AASHTO T 152 Sec. Annex A1.2
  Calibration Interval Occasionally
     
  Calibration Item Effective volume of the vessel
  Calibration Procedure AASHTO T 152 Sec. Annex A1.4
  Calibration Interval Occasionally
 

  Item Coil spring or other device
  Description Used to hold the calibration cylinder in place
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Sieves
  Description 1.5" pan sieve
     
  Calibration Item Size of openings and wire diameter
  Calibration Procedure AASHTO T 152 Sec. 7.1
AASHTO M 92 (ASTM E11, Annex A1) via AASHTO T 141
  Calibration Interval None stated
 

  Item Tamping Rod
  Description 5/8" diameter straight steel rod, with hemispherical tip, not less than approximately 16" in length
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Mallet
  Description Rubber or rawhide, 1.25 + 0.50 lb., or 2.25 + 0.50 lb.
     
  Calibration Item Mass
  Calibration Procedure Weigh with a balance
  Calibration Interval None stated
 

  Item Strike-off bar or plate
  Description For removing excess concrete from the top of the mold, flat rectangular steel or acrylic plate of appropriate dimensions for each material
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Funnel
  Description Used to aid the pouring of water in to the meter
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Measure for water
  Description A container to hold the water that is poured in to the meter
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Miscellaneous Tools
  Description Shovel, scoop, trowel
     
  Calibration Item See the general statement above.
  Calibration Procedure  
  Calibration Interval  
 

  Item Internal Vibrator
  Description 7000 vibrations per minute, or greater, while in use and of appropriate size .
     
  Calibration Item Frequency
  Calibration Procedure AASHTO T 152 Sec. 4.15
With a vibrating reed tachometer, per AASHTO T 126 Sec. 3.6
  Calibration Interval None stated
 

  Item External Vibrator
  Description 3600 vibrations per minute, or greater..
     
  Calibration Item Frequency
  Calibration Procedure AASHTO T 152 Sec. 4.15
With a vibrating reed tachometer, per AASHTO T 126 Sec. 3.6
  Calibration Interval None stated