Equipment Checklist for AASHTO T 11-91
Materials Finer than .075-mm (No. 200) Sieve
in Mineral Aggregates by Washing

 
All equipment shall be inspected for compliance with appropriate dimension s, as stated in each standard, as well as to determine if it is in satisfactory operation condition (i.e. clean, tight and serviceable). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.

 

 

  Item Sieves
  Description Nest: Upper sieve opening size ranging from 2.36 mm (No. 8) to 1.18 mm (no. 16), Lower sieve opening size 0.075 mm (No. 200)
     
  Calibration Item Opening size
  Calibration Procedure AASHTO M 92
  Calibration Interval Not stated
 

  Item Balance
  Description Conforms to AASHTO M-231, accurate to 0.1 percent of sample mass
     
  Calibration Item Accuracy and condition
  Calibration Procedure AASHTO M 231, Sec. 4.2 and 4.3
  Calibration Interval Not stated
 

  Item Oven
  Description Thermostatically controlled, forced draft, 110C + 5
     
  Calibration Item Temperature
  Calibration Procedure AASHTO T 265, sec. 3.1
  Calibration Interval Not stated