Equipment Checklist for AASHTO T 104
Soundness of Aggregate by Use of Sodium Sulfate or Magnesium Sulfate

 
All equipment shall be inspected for compliance with appropriate dimensions, as stated in each standard, as well as to determine if it is in satisfactory operating condition (i.e.; adherence of mortar, deformed molds, etc.). Inspection shall occur at least annually, or whenever there is reason to suspect nonconformance with the standard.

 

 

  Item Sieves, #4, #5, #8, #16, #50, #100
  Description Wire cloth mounted in a frame
     
  Calibration Item Check openings
  Calibration Procedure

See Annex A1.2, of AASHTO M 92

  Calibration Interval 6 months
 

  Item Apparatus for immersing samples in solution (optional)
  Description Check physical condition
     
  Calibration Item Check physical condition
  Calibration Procedure Visual
  Calibration Interval 12 months
 

  Item Balance
  Description Readable to 0.1% of sample mass or better
     
  Calibration Item Verify mass readings
  Calibration Procedure AASHTO M 231
  Calibration Interval 12 months
 

  Item Thermometer
  Description Readable to 0.1 C
     
  Calibration Item Temperature Readings
  Calibration Procedure ASTM C 1084 via AASHTO T 23, Sec. 7.3
  Calibration Interval See ASTM C 1084
 

  Item Oven
  Description Must be maintained at 110 + 5 C
     
  Calibration Item Verify temperature
  Calibration Procedure Verify with calibrated thermometer
  Calibration Interval 4 months
 

  Item Temperature Regulation
  Description Suitable means for regulating temperature of samples during immersion in solutions
     
  Calibration Item Verify temperature
  Calibration Procedure Verify with calibrated thermometer
  Calibration Interval See ASTM C 1064
 

  Item Hydrometer
  Description Conforming to Sec. 5-11 in Specification E 100
     
  Calibration Item Capable of measuring liquid within + 0.01
  Calibration Procedure Verify measurement
  Calibration Interval 12 months
 

  Item Containers
  Description Must permit free access of solution to sample and drainage of solution from sample without loss of aggregate
     
  Calibration Item Access and drainage of solution
  Calibration Procedure Visual
  Calibration Interval 12 months
 

   

The following special solutions are required to perform this test method:

Sodium sulfate and Magnesium sulfate solution